Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Barnant 115: (1 year) CAL VER /5520 DATE: 27-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 25 NUMBER OF LINES: 110 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # BARNANT 115 Manual No. A-1925-27 # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- P J F W 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Warm-up time: NONE 1.005 DISP Temperature: 18 - 28 degC 1.005 DISP Humidity: <80% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 HEAD THERMOCOUPLE 1.008 MEM2 Type the letter of the Thermocouple type you are using. 1.009 JMPL PROCEED ZCMPI(MEM2, "K") 1.010 JMPL PROCEED ZCMPI(MEM2, "J") 1.011 JMPL PROCEED ZCMPI(MEM2, "T") 1.012 JMPL PROCEED ZCMPI(MEM2, "E") 1.013 DISP The selected type of thermocouple is NOT available. 1.014 JMP 1.008 1.015 LABEL PROCEED 1.016 DISP Connect the 5520A TC Terminals to the INPUT JACK 1.016 DISP (T2 connector, right side), using a [MEM2] type 1.016 DISP thermocouple wire and connectors. Observe polarity. 1.017 DISP Press the ON/OFF button to turn the UUT ON. 1.017 DISP Press the J.T.E.K pushbutton to select [MEM2] type. 1.017 DISP Press the 1/.1deg pushbutton to select .1deg. 1.017 DISP Press the DEG C/F pushbutton to select F. 1.018 JMPL K_TYPE ZCMPI(MEM2, "K") 1.019 JMPL J_TYPE ZCMPI(MEM2, "J") 1.020 JMPL T_TYPE ZCMPI(MEM2, "T") 1.021 JMPL E_TYPE ZCMPI(MEM2, "E") 1.022 LABEL K_TYPE 1.023 HEAD {K TYPE THERMOCOUPLE} 1.024 5520 -300.0degF 0.1% 0.7U _K TC 2W 2.001 5520 0.0degF 0.1% 0.7U _K TC 2W 3.001 5520 2500.0degF 0.1% 0.7U _K TC 2W 4.001 DISP Select degC. 4.002 5520 -200.0degC 0.1% 0.4U _K TC 2W 5.001 5520 32.0degC 0.1% 0.4U _K TC 2W 6.001 5520 1370.0degC 0.1% 0.4U _K TC 2W 7.001 JMP 25.001 7.002 LABEL J_TYPE 7.003 HEAD {J TYPE THERMOCOUPLE} 7.004 5520 -300.0degF 0.1% 0.4U _J TC 2W 8.001 5520 0.0degF 0.1% 0.4U _J TC 2W 9.001 5520 1800.0degF 0.1% 0.4U _J TC 2W 10.001 DISP Select degC. 10.002 5520 -100.0degC 0.1% 0.7U _J TC 2W 11.001 5520 32.0degC 0.1% 0.7U _J TC 2W 12.001 5520 900.0degC 0.1% 0.7U _J TC 2W 13.001 JMP 25.001 13.002 LABEL T_TYPE 13.003 HEAD {T TYPE THERMOCOUPLE} 13.004 5520 -400.0degF 0.1% 0.7U _T TC 2W 14.001 5520 0.0degF 0.1% 0.7U _T TC 2W 15.001 5520 750.0degF 0.1% 0.7U _T TC 2W 16.001 DISP Select degC. 16.002 5520 -200.0degC 0.1% 0.4U _T TC 2W 17.001 5520 32.0degC 0.1% 0.4U _T TC 2W 18.001 5520 390.0degC 0.1% 0.4U _T TC 2W 19.001 JMP 25.001 19.002 LABEL E_TYPE 19.003 HEAD {E TYPE THERMOCOUPLE} 19.004 5520 -300.0degF 0.1% 0.7U _E TC 2W 20.001 5520 0.0degF 0.1% 0.7U _E TC 2W 21.001 5520 1700.0degF 0.1% 0.7U _E TC 2W 22.001 DISP Select degC. 22.002 5520 -200.0degC 0.1% 0.4U _E TC 2W 23.001 5520 32.0degC 0.1% 0.4U _E TC 2W 24.001 5520 900.0degC 0.1% 0.4U _E TC 2W 25.001 END